Process temperature
-50 … 230 °C , without cooling neck @ Tamb = 20 °C
-50 … 400 °C , with cooling neck @ Tamb = 20 °C
Pt100 accuracy class (EN 60751)
B (± 0.3 °C at 0 °C)
± (0.3 + 0.005 × t)°C
A (± 0.15 °C at 0 °C)
± (0.15 + 0.002 × t)°C
1/3 B (± 0.1 °C at 0 °C)
± 1/3 × (0.3 + 0.005 × t)°C
1/6 B (± 0.05 °C at 0 °C)
± 1/6 × (0.3 + 0.005 × t)°C
Response time, T50
< 1.5 s , Ø4 mm
< 6.1 s , Ø6 mm
< 7.6 s , Ø8 mm
Process connection
Sensor length
20 … 3000 mm
Sensor tube material
AISI 316L (1.4404)
Surface roughness wetted parts
Ra ≤ 0.8 µm
Ambient conditions
Degree of protection (EN 60529)
IP 67
IP 69K , with appropriate cable
Humidity
< 98 % RH , condensing
Vibration (sinusoidal) (EN 60068-2-6)
1.6 mm p-p (2 … 25 Hz), 4 g (25 … 100 Hz), 1 octave / min.
Output signal
Without transmitter
Pt100, 4-wire
With transmitter
4 … 20 mA , 2-wire
4 … 20 mA , + HART®
Compliance and approvals
EMC
EN 61000-6-2
EN 61000-6-3
Pressure directive
97/23/EG
Hygiene
3-A (74-06)
FDA (21 CFR 177.2415