Process temperature
-50 … 400 °C
-50 … 600 °C , optional
Max. flow velocity
40 m/s , gases
5 m/s , liquids
Pt100 accuracy class (EN 60751)
B (± 0.3 °C at 0 °C)
± (0.3 + 0.005 × t)°C
A (± 0.15 °C at 0 °C)
± (0.15 + 0.002 × t)°C
1/3 B (± 0.1 °C at 0 °C)
± 1/3 × (0.3 + 0.005 × t)°C
1/6 B (± 0.05 °C at 0 °C)
± 1/6 × (0.3 + 0.005 × t)°C
Pt1000 accuracy class (EN 60751)
B (± 0.3 °C at 0 °C)
1/3 B (± 0.1 °C at 0 °C)
Response time, T50
≤ 1.5 s , Ø4 mm
≤ 6.1 s , Ø6 mm
≤ 7.6 s , Ø8 mm
≤ 11.1 s , Ø10 mm
Ambient conditions
Degree of protection (EN 60529)
IP 65
Humidity
≤ 100 % RH , condensing
Long-term test (transmitter)
IEC 770 6.3.2
Vibration (sinusoidal) (EN 60068-2-6)
1.6 mm p-p (2 … 25 Hz), 4 g (25 … 100 Hz), 1 octave / min.
GL, test 2
Output signal
Without transmitter
Pt100, 2-wire
Pt100, 4-wire
With transmitter
4 … 20 mA , 2-wire